HP Agilent 1100 Series G1364A AFC-automatic fraction collector

Type Specification
Trigger modes Time slices,
Peak (threshold, up- / downslope),
Timetable (combination of time intervals and peak) and Manual trigger (supported only with G1323B Control Module)
Agilent 1200 Series DAD/MWD detectors (G1315A/B/C, G1365 A/B/C), the Agilent 1200 Series fluorescence detector and the Agilent G1946C/D, G1956A/B LC-MSD are fully supported other detectors can be used but are not supported for fraction collection
operating modes Discrete fractions: default mode for all vessels. The flow is diverted to waste, while moving from one vessel position to the next vessel positionContinuous flow: optional, available only when using well plates. It is possible to move from one well plate position to the next one without diverting the flow into the well plate to waste
Fraction capacities
and trays 4 x well-plates full tray (MTP)*
(for use with deep well plates, only)
2 × well-plates std. tray (MTP) (for use with deep well plates, only)
+ 10 × 2 ml vials* (+ 1 half tray)
100 x 2 ml in std. tray (+ 1 half tray)*
3 x 40 x 2 ml in half tray*
3 x 15 x 6 ml in half tray*
Full tray with 40 test tubes (30 mm OD, max. height 100 mm, ~45 ml / tube)
Full tray with 60 test tubes (25 mm OD, max. height 100 mm, ~25 ml / tube)
Full tray with 126 test tubes (16 mm OD, max. height 100 mm, ~12 ml /
tube)
Full tray with 215 test tubes (12 mm OD, max. height 100 mm, ~7 ml / tube)
Installed trays are automatically detected and identified.
For the with uncapped vials, tests tubes and well plates, only!
test tube / plate sizes Minimum 48 mm to 100 mm maximum
Maximum tube
Volume ca. 45 ml
Maximum flow rate 100 ml / min (depending on viscosity and generated back pressure, max.
6 bar at the diverter valve)
Delay volumes [μl] Fraction collector inlet to diverter valve: ~500 (typical, depends on length of
the tubing)
Diverter valve: ~15
Diverter valve to needle: ~110
Needle: ~5
Delay calibration Sensor Single wavelength absorbance detector working at 654 nm, consisting of a LED and a photo diode
Diverter valve 3/2 Diverter valve with low internal volume (15 μl), switching time < 100 ms, maximum operating pressure 6 bar
Cooling Optional (with additional G1330B), performance depending on ambient conditions and the volume of collected fractions
maximum capacity 3 fraction collectors in parallel plus one recovery fraction collector connected via 12-Position, 13-Port Selector valve (PN G1160A)
GLP features Early maintenance feedback (EMF), electronic records of maintenance and
Errors
Interfaces Controller-area network (CAN).
optional; LAN or external contacts interfaceRS232C,APG-remote (for remote start / stop signals to / from other modules) Interface to G1330A Thermostat CAN-DC-out for operation of Agilent approved external devices like valves
Safety features Leak detection and safe leak handling, error detection and display, exhaust fan for fume extraction of hazardous vapors

This refurbished HP / Agilent 1100 Series G1364A AFC-automatic fraction collector is in great condition.