Air Shields TI 100 Infant Incubator
AMERICA North (USA-Canada-Mexico)
Condition Refurbished
FEATURES
Intended for transport of high risk, premature, low birth-weight, or critically ill newborns.
It provides a means to control air temperature, oxygen concentrations, and relative humidity.
A double wall hood permits full visibility and provides an effective thermal barrier from the environment.
SPECIFICATIONS
Dimensions (with stand):
Length: 40.25 in. (102 cm)
Width: 22.25 in. (56.5 cm)
Height: 46.75 in. (118.7 cm) maximum, 34.75 in. (88.3 cm) minimum
Power Requirements:
AC: 110.120 VAC 50/60/400 Hz, sine or square wave 170/200 Watts and 225/270 Watts with batteries charging.
DC: 12 VDC, 115 Watts
Internal Battery:
Type: gel type, sealed and rechargeable.
Voltage: 12 VDC nominal
Quantity: 2
Charging Time: two batteries take 20 minutes to recharge from fully discharged.
Life expectancy: minimum 200 full charge/discharge cycles.
Chassis Leakage Current: 100µA or less
Alarms:
High Temp: Actuates at displayed incubator temperature of 39º ± 0.5 ºC.
Sensor: Actuates if High Temp alarm sensor is electrically open or shorted.
Power Failure: Actuates if main AC power fails and no DC back-up power is present.
Heater Temp: Actuates if heater temperature exceeds 77 ºC nominal.
Low DC: Actuates if incubator DC voltage has fallen below 10.5 VDC nominal.
Temperature:
Set Point Range: 21.5 ºC ± 1.0 ºC to 38 ºC
Rise Time: 30 minutes
Variability: 0.2 ºC
Overshoot: 1.5 ºC maximum
Uniformity: 1 ºC
Correlation of Displayed Incubator to Actual Incubator Temperature at Temperature Equilibrium: 1 ºC
Correlation of Displayed Incubator Temperature to Set Point at Temperature Equilibrium: ± 0.5 ºC
Observation Light Output: 10-foot candles at mattress level