Brand: Oricare
Oricare Ventilator
Pneumatically driven and electronically controlled
ventilation
Suitable for children and adults
Easy control of user interface by touch screen &
navigator wheel
Sensitive adjustable flow trigger and pressure trigger
Lung mechanics monitoring and flow loops
15”TFT LCD color screen
Synchronized nebulizer
Inspiratory and expiratory hold
Manual inspiration
Smart suction
Built-in battery (≥1hour) or optional extended
batteries (≥2hours operation)
Easy to position ergonomic Trolley available with
lockable drawer and auxiliary AC outlets
Easily removeable and sterilizeable exhalation
valve module

Modes of Ventilation

Pressure Controlled (PCV), Assist/Control
Pressure Regulation Volume Controlled (PRVC),
Assist/Control
Synchronized Intermittent Mandatory
Ventilation, SIMV(VCV) + PSV, SIMV(PCV) + PSV,
SIMV(PRVC) + PSV
Spontaneous Ventilation (SPONT/CPAP) + PSV
Bi Level Ventilation (BIVENT/APRV) + PSV
Non-invasive Ventilation (NIV/CPAP + PSV,
NIV/PCV)
Volume Controlled (VCV), Assist/Control

Parameters and Ranges

Tidal volume: 20-2500ml
Respiratory rate: 1-120bpm
Inspiration time:0.2-9s
Inspiratory pause time: 0-4s
Tslope : 0-2s
FiO2: 21%~100%
PEEP: 0-50cmH2O
CPAP: 2-20cmH2O
Psupport: 0-90cmH2O
Pinsp: 5-90cmH2O – USA
0-100cmH2O – International
Esens: 5%-80%
Flow patterns: square, decelerating
Inspiratory hold
Expiratory hold
Nebulizer: 30 or 45 minutes
Auto tube compensation
Compliance compensation
Graphical trends
Nurse call

Monitoring

Pressure values: Pmin, Pplat, Pmean, Ppeak, PEEP
Volume/Flow values: VTI, VTE, MV, MVspont, Leak%
Loops: Flow – Pressure, Pressure – Volume,
Flow – Volume
FiO2
Optional etCO2

Alarms

MV high, MV low, Circuit disconnect, Paw high, Paw low, Continuous airway pressure high, VTi high, fspont high, Tapnea, FiO2 high, FiO2 low, etCO2 high, etCO2 low, Power supply failure, Battery low, Battery exhausted, Air supply failure, O2 supply failure, PEEP low.

Operating Conditions

Gas supply: O2, Air, 200kPa -650kPa (29-94PSI)
Power supply: 100-240V, 50-60Hz
Temperature: 10-40 (Operation)
-20-60℃(Storage)
Relative humidity: 5-95%, non-condensing
(Operation)
≤95%, non-condensing (Storage)
Atmospheric pressure: 500hPa-1060hPa
(Operation&Storage)
Communication interfaces: RS232, VGA, USB, Rj45
Dimensions (H×W×D): 624x774x1410mm

Standards

UL/IEC 60601-1, IEC 60601-1-2, IEC 60601-2-12, EN 794-1