Product Specifications
Operating Principle:
Linear peristalsis
Dimensions:
10.6″ x 11.9″ x 7.2″ (26.9 cm x 30.2 cm x 18.3 cm)
Weight:
18.5 lbs (8.4 kg)
Power Requirements:
90 – 130 (120 nominal) VAC, 0.4A, 50/60 Hz, 3 Wire, single phase
Battery Operation:
With a new, fully charged battery, approximately 5 hours with one channel operating at 125
ml/hr, or approximately 4 hours with two channels operating at 125 ml/hr.
Electrical Leakage:
Less than 100 microamps
Operating Range
Controller Mode:
Actual: 1-999 ml/hr (in 1 ml/hr increments) Recommended: 1-500 ml/hr
Pump Mode:
1-999 ml/hr (in 1 ml/hr increments)
KVO (Keep Vein Open) rate:
1 ml/hr
Volume to be Infused range:
1-9999 ml, ALL
Occlusion Pressure
Controller Mode:
Container height
Pump Mode:
10 psi for rates above 30 ml/hr. For rates Below 30 ml/hr, the occlusion pressure is Rate dependent.
Advisories:
Secondary, Primary, Infusion Complete – KVO, Empty Container – KVO, Low Battery, Maintenance, Pause, Low Flow, Total Vol Infused, and Sec Vol Infused.
Prompts:
Select PC Mode, Set Rate, Set VTBI, Press Start, and Access Channel
Alarms:
Check ECD, Check IV Set, Close Door, Air-in-line, Occluded Patient Side, Occluded Fluid Side, Occluded, Partial Occlusion Fluid Side, Flow-Stop Open/Close Door, KVO Comm Tout, and KVO