ASP Sterrad 100 NX
EUROPE (Central and Eastern)
Plasma Sterilizer
Chamber Capacity: 152 Liters (rectangular), two-tiered shelving.
Cycle Times (approx.):
EXPRESS: 24 minutes
FLEX: 42 minutes
STANDARD: 47 minutes
DUO: 60 minutes
Dimensions:
Width: 775 mm
Height: 1800 mm
Depth (Single Door): 1055 mm
Depth (Double Door): 1095 mm
Sterilant Delivery: Cassettes with automatic expiration date detection.
Interface: Projected capacitive touchscreen (800 x 600 pixels).
Weight: 382 kg (single door), 408 kg (double door).
Safety/Process Monitoring: ALLClear™ Technology for detecting load/system issues, barcode reader, USB data export.
Configurations: Available in single and double doors.
Cycle Suitability
STANDARD Cycle: General metal/non-metal devices, cameras,
light cords, rigid/semi-rigid endoscopes.
FLEX Cycle: Single-channel flexible endoscopes.
EXPRESS Cycle: Fast turnaround for da Vinci endoscopes, rigid scopes,
non-lumened items.
DUO Cycle: Specialized for complex, flexible scopes.
Installation Requirements
Electrical: 208 VAC, 60 Hz, 3-phase wye connection (US/Canada); 380-415 VAC, 50/60 Hz
(International).
Environment: Dry, well-ventilated area with consistent ambient temperature