Biorad CFX Opus 96
AMERICA North (USA-Canada-Mexico)
Real Time PCR
Maximum ramp rate 5°C/sec
Average ramp rate 3.3°C/sec
Heating and cooling method Peltier
Lid 30–110 °C
Temperature Range 4–100°C
Accuracy ±0.2°C of programmed target at 90°C
Uniformity ±0.3°Cwell-to-well after 10 sec of arrival at target
Gradient
Operational range 30–100°C
Programmable span 1–24 °C
Optical Detection
Excitation 6 filtered LEDs
Detection 6 filtered photodiodes
Range of excitation wavelengths, nm 450–684
Range of emission wavelengths, nm 515–730
Sensitivity Detects 1 copy of target sequence in human genomic DNA
Dynamic range 10 orders of magnitude
Software Operating systems Windows 10
Touch Screen Adjustable from 15°–22°
Sample capacity 96-well
Sample size, µl 1–50 (10–50 recommended)
Communication interface USB 2.0
Dimensions 13x26x14”
Weight 48lbs
Power 100-240v, 50/60hz, 850W