Evone Medical Mechanical Ventilator
EUROPE (Western and Northern)
VENTINOVA Medical EVONE
Features
Controlled ventilation throughout the respiratory cycle
Improved lung recruitment and aeration
Reduction of atelectasis and improvement of gas exchange
Accurate calculation of energy dissipation for protective ventilation
Adaptation to narrow airways via Tritube®
Optimisation of single-lung ventilation
Jet Ventilation mode to facilitate weaning and spontaneous breathing
Continuous monitoring of respiratory parameters
Technical Details
Flow-Controlled Ventilation (FCV®) technology with
active expiratory control
Compatible with endotracheal tubes from
2 mm to 10 mm internal diameter
Direct, real-time measurement of intratracheal pressures
Main settings: inspiratory flow rate, I:E ratio,
peak pressure (Peak), expiratory pressure (EEP)
Typical flow rates between 8 and 16 L/min
High-frequency jet ventilation mode
with maximum pressure of 1.5 bar
For use on patients ≥ 40 kg in the operating theatre and
intensive care
Compatible with Tritube®, CTA and double-lumen tubes
Compatible Accessories
Tritube®
Adapter for conventional endotracheal tube (CTA)
Tritube® patient kit for 20 patients (18055)
CTA patient kit for 20 patients (18075)
Airway adapter (6125)
Cartridge (6115)
Heat and moisture exchanger (HME) filter (12012)
Breathing circuit (6120)
Transport trolley (18030)