Hewlett Packard
AMERICA North (USA-Canada-Mexico)
Hewlett-Packard M1722A CodeMaster XL+ Defibrillator, Monitor, Recorder with Pacing and Sync
Includes:
3 Lead Patient ECG Cable
No Paddles
Specifications-
Defibrillator:
Waveform: Damped sinusoidal
Output energy (delivered): 2, 3, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 300, and 360 joules.
Charge control: Push-button on apex paddle and on front panel
Charge time: Less than 5 seconds to 360 joules
Charge done indicator: Display indicates available energy, charge done tone.
Synchronizer: SYNC message appears on the monitor and is annoted periodically on the recorder while in synchronous mode. An audible beep sounds with each detected R-wave, while a marker on the monitor and sync mark on the recorder strip indicate the discharge point.
Monitor:
Input: ECG activity may be viewed through the paddles or patient cable
Lead fault: LEADS OFF message and dashed baseline appear on the monitor if a lead becomes disconnected. Common mode rejection: greater than 100dB measured as per AAMI standards for cardiac monitors.
Heart rate display: Digital read out on monitor from 20 to 280 bpm
Recorder:
Recorder Mode: Automatically documents events and ECG during defibrillation episodes. The recorder can be configured to run in either real time or with a 6-second delay.