Hewlett-Packard Model: Codemaster XL

Features:
Defibrillator with ECG and recorder
Provides 1-2-3 operation
Built-in pediatric paddles under slide-off adult paddles
Preset heart rate alarms
High-capacity battery
Built-in diagnostics

Specifications:
Dimensions: 7.9" x 11.8" x 15.6" (20cm x 30cm x 39.7cm)
Weight: 24 lbs. / 10.9kg

Battery:
Type: Rechargeable, sealed lead-acid battery. 4 Ah, 12 V nominal.
Charge Time: 2 hours to 90% of full capacity. 18 hours to 100% capacity. Repeated charging to less than 100% will reduce the useful life of the battery.
Capacity: 2.5 hours monitoring or 50 full-energy discharges or 1 hour monitoring and recording.
AC input: 100 to 230 VAC +/- 15%, 50/60 Hz.
Battery Indicators: Illuminated LED indicates battery is charging. Low Battery message appears on the monitor.

Defibrillator
Waveform: damped sinusoidal.
Output energy (delivered): 2,3,5,7,10,20,30,50,70,100,150,200,300, and 360 joules.
Charge control: push-button on apex paddle and on front panel.
Charge time: less than 5 seconds to 360 joules with a functional battery installed.
Charge done indicator: display indicates available energy, charge-done light on apex paddle, charge done tone.
Synchronizer: SYNC message appears on the monitor and is
annoted periodically on the recorder while in synchronous mode.
An audible beep sounds with each detected R-wave, while a
marker on the monitor and sync mark on the recorder strip indicate the
discharge point.
Paddles: standard paddles are anterior / anterior. Adult electrodes
(83 cm sq) slide off to expose pediatric electrodes (21 cm sq).
Paddle cord is 10 feet (3m). Full range of internal paddles are available.

Monitor
Input: ECG activity may be viewed through the paddles or patient cable.
Lead fault: LEADS OFF message and dashed baseline appear on the
monitor if a lead becomes disconnected. Common mode rejection: greater
than 100dB measured as per AAMI standards for cardiac monitors.
Heart rate display: digital read out on monitor from 20 to 280 bpm.

Recorder
Annotations: time, date, heart rate, event markers, ECG mode, and selected energy.
Recorder Mode: automatically documents events and ECG during defibrillation episodes. The recorder can be configured to run in either real time or with a 6-second delay.
Frequency response: .5 to 40Hz monitor bandwidth. .05 to 150 Hz diagnostic bandwidth is selectable.