Menu

OEC 9800

Ref : 2153837-14
Condition : Used
Manufacturer : OEC
Model : 9800
Year(s) : 2001
Quantity : 1
Location : Seller or machines location:
AMERICA North (USA-Canada-Mexico)

C ARM
2010 TUBE
REV 30 SW
VASCULAR
Specialty Packages Available
• Vascular 8 Generator
• 60 kHz High frequency
• 15 kW full-wave
• Up to 120 kVp
• Up to 75 mA for radiographic film exposures
• Continuous fluoro boost mode up to 20 mA
• Pulsed fluoro mode
• Pulsed fluorography mode up to 40 mA (optional)
• Full power from outlet:
– 50 Hz or 60 Hz
– 100V-15A to 240V-10A
• Patented energy buffer design
X-Ray Tube-2010
• Rotating anode X-ray tube
• Focal spots: 0.3 mm and 0.6 mm
• Anode heat capacity: 300,000 HU
• Anode cooling rate: 60,000 HU/min
• Housing heat capacity: 1,600,000 HU
• Anode diameter: 3.1 in
• Anode angle: 10°
• X-ray tube cooling fan (Vascular,Neurovascular & Cardiac units)
Collimator
• Iris collimation with dual opposing semitransparent leaf shutters allow elliptical,round and rectangular collimation
• Minimum beam size < 5 X 5 cm2
• Iris and shutters are continuously variable
• All functions remotely controlled from the C-Arm control panel Fluoroscopy Mode
• Focal spot: 0.3 mm
• kVp range: 40 – 120 kVp
• mA range: 0.2 – 10 mA normal mode
• mA range: 0.2 – 20 mA continuous fluoro boost mode
• Auto and manual modes
• Continuous, one-shot or pulsed operation
• ABS varies mA, kVp and camera gain
• User specific ABS tables Pulsed Fluoroscopy Mode
• Focal spot: 0.3 mm
• kVp range: 40 – 120 kVp
• mA range: 0.2 – 10 mA
• Pulse rate: 1, 2, 4, or 8 pulses per second
• Pulse width: 30 or 50 milliseconds
• Computer controlled ABS, mA, kVp and camera gain Pulsed Fluorography Mode (optional)
• Focal spot: 0.3 mm
• kVp range: 40 – 120 kVp
• mA range: Up to 40 mA
• Pulse rate: 1, 2, 4, or 8 pulses per second
• Pulse width: 30 or 50 milliseconds
• Computer controlled camera iris, mA, kVp, & camera gain Digital Spot Mode
• Focal spot: 0.3 mm
• kVp range: 40 – 120 kVp
• mA range: 1.0 – 75 mA
• Automatically terminates exposure and stores enhanced image to storage archival disk Radiographic Mode
• Focal spot: 0.3 mm or 0.6 mm
• Focal spot automatically selected
• kVp range: 50 – 120 kVp
• 0.3 mm – mAs range: 1 – 100 mAs
• 0.6 mm – mAs range: 110 – 300 mAs

Video Imaging System Image Intensifier 12 in Image Intensifier
• Tri-mode 12 in/9 in/6 in image intensifier
– 12 in.: 1.7 lp/mm
– 9 in.: 2.3 lp/mm
– 6 in.: 3.0 lp/mm
• DQE: 65% (typical)
Video Camera
• High resolution CCD camera
• 1,004 x 1,004 pixels
• Full frame capture
• 360° motorized rotation
• On-screen orientation indicator (realtime feedback without fluoro)
• Left-right image reversal
• Top-bottom image reversal
• Negative mode
• Computer controlled features:
– Gain
– Blanking
– Camera iris

Video Monitors :
• Dual 16 in, black etched, anti-glare, anti-static
• Touch Screen Display
• Ambient room light sensor
• Multi-LED X-ray “ON” indicator
• Remote window/level controls

Hardcopy Options :
• Medical Printer
• DICOM
• Image Capture Device

C-Arm Dimensions :
• Height: 69 / 87 in (1734 / 2191 mm)
• Width: 33 in (838 mm)
• Length: 79 in (1994 mm)
• Weight: 720 lbs

Workstation Dimensions :
• Height: 64 in (1632 mm)
• Width: 27 in (692 mm)
• Depth: 27 in (692 mm)
• Weight: 430 lbs.

Other machines similar to OEC 9800

1
Location : EUROPE (Western and Northern)
Year(s) : 2003
1
Location : AMERICA North (USA-Canada-Mexico)
1
Location : AMERICA North (USA-Canada-Mexico)
Year(s) : 2012
1
Location : AMERICA North (USA-Canada-Mexico)
Year(s) : 2000