Perkin Elmer Flexar FX-15 UHPLC

This refurbished Perkin Elmer Flexar FX-15 UHPLC is like new. It includes Flexar Solvent Manager 3-CH Degasser. It has been tested at SpectraLab Scientific.

Features

Dual-reciprocating piston 18,000 psi pump provides as much as 10x or more improvement in productivity
Integrated piston wash function keeps pump pistons clean, even with buffers
Synchronized Injection for ultimate precision. High-speed FXUV/Vis (100 pt/sec) with high-efficiency 2.4 μL flow cell
View pump pressure profiles real-time with Chromera CDS
Analyst versions supported: version 1.4.2 and higher
Supports 21 CFR Part 11 Compliance
Up to 181 gradient profiles, including step, linear and both convex and concave curves
Up to 19 gradient steps plus an equilibration step
Two independent timed events, including a third timed event that triggers both simultaneously
Soft Ramp option for automatically ramping the flow rate when starting and stopping the pump, placing less stress on
UHPLC systems and prolonging column life
FX-15 supports GradientLock, timing each injection with the pump's piston position, driving improved retention time repeatability

Flexar Solvent Manager 3-CH Degasser specification

Specification
Description
Channels
3-channel version or 5-channel version
Degassing Process
Gas permeation through a fluoropolymer membrane
Maximum Recommended Flow Rate
3.0 mL/min.
Pressure Drop
1.37 mm Hg/mL/min.
Degassing Capacity
~25% dissolved gases remaining in 60:40 MeOH/Water mixture at 1 mL/min.
Dead Volume
~480 microliters per channel for standard channel
Materials contacting solvents
PEEK, Glass-filled PTFE, Teflon AF
Power:

Power requirement
100 to 240 VAC (±10%), 1A, 47 to 63 Hz
Power Supply Adapter Plugs
4 supplied with power supply, interchangeable: North America/Japan, U.K., Continental Europe, Australia
Installation Over-Voltage Category
II
Validation Output:

Signal
5 mVDC / 1 mm Hg absolute from 20 to 800 mm Hg (0.100 VDC at 20 mm Hg; 4.000 VDC at 800 mm Hg)
Accuracy
±1.0% of reading ±0.010 VDC from 20 to 800 mm Hg
Operating Conditions:

Ambient Temperature
10 to 35 ºC
Ambient Relative Humidity (RH)
20 to 80 % RH (without condensation)
Altitude
0 to 2000 Meters
Indoor vs. Outdoor Use
Indoor
Pollution Degree
2
Storage Conditions:

Ambient temperature
-20 to +60 ºC
Ambient Relative Humidity
20 to 80% RH (without condensation)
Altitude
0 to 12000 M