Philips V60 Ventilator
AMERICA North (USA-Canada-Mexico)
Pressure Range:
IPAP (Inspiratory Positive Airway Pressure): 4 to 40 cmH₂O.
EPAP (Expiratory Positive Airway Pressure): 4 to 25 cmH₂O.
CPAP: 4 to 25 cmH₂O.
Target Tidal Volume (AVAPS): 200 mL to 2000 mL.
Respiratory Rate: 4 to 60 breaths per minute (BPM).
Inspiratory Time: 0.30 to 3.00 seconds.
Oxygen Concentration (): 21% to 100%.
The Philips V60 supports various breathing modes for clinical flexibility:
CPAP (Continuous Positive Airway Pressure) with the C-Flex feature for
expiratory comfort.
S/T (Spontaneous with Timed Backup): Provides breath support when the
patient begins to inhale or automatically if the respiratory rate decreases.
PCV (Pressure Control Ventilation).
AVAPS (Average Volume Assured Pressure Support): Ensures target tidal volume
by automatically adjusting pressure.
PPV (Proportional Pressure Ventilation): An additional option that
delivers pressure according to the patient's respiratory effort.