6 Welch Allyn LXi Spot Vital Signs Monitor 45NT0
AMERICA North (USA-Canada-Mexico)
Welch Allyn LXi Spot Vital Signs Monitor 45NT0
The Spot Vital Signs LXi automatically measures systolic and diastolic pressure (excluding neonates), pulse rate, temperature (oral, adult axillary, pediatric axillary, rectal, and ear), and pulse oximetry (SpO2) as well as calculates Mean Arterial Pressure (MAP).
Furthermore, Spot Vital Signs LXi allows the entry of height, weight, respiration rate, and pain level. Spot Vital Signs LXi also calculates Body Mass Index (BMI) following height and weight entry.
 
FEATURES
 
SureBP with pulse rate and MAP
SureTemp™® Plus or Braun ThermoScan® PRO 4000 thermometry
Masimo® or Nellcor pulse oximetry
Manual weight, height, respiration ® rate, and pain level entry
Compatible with select digital scales
Calculates Body Mass Index
50-reading memory
Connectivity-ready
Fast, accurate, flexible
Easy to use
 
 
SPECIFICATIONS
 
Display: 16 character liquid crystal display 
Physical: H: 10.6 in., W: 7.5 in., D: 5.25 in., Weight: Approx. 7.5 lbs.
 
Non-Invasive Blood Pressure (NIBP)
Cuff Pressure Range: 0 to 300 mmHg, 0 to 40 kPa 
Time: 15 seconds,
Overpressure Cutoff: 315 to 15 mmHg (42 to 1.9 kPa) 
Pulse Rate Range: 35 to 199 bpm 
Pulse Rate Accuracy: ± 5%
 
SpO2: 
*Performance Measurement: 1% to 100% 
*Accuracy: Adult/Pediatric 70-100% ±2 digits, Neonate 70-100% ±3 digits 
*Pulse Rate Range: 20 to 255 bpm 
*Pulse Rate Accuracy: 25 to 240 bpm ±3 digits
 
Temperature: 
Measurement Range: 80F to 110F (26.7C to 43.3C) 
Calibration Accuracy: ± 0.2F (±0.1C);
Determination Time (approximate): Predictive 
-Oral, 4-6 seconds 
-Pediatric Axillary, 10-13 seconds 
-Adult Axillary, 12-15 seconds 
-Rectal, 10-13 seconds Monitored