Menu

Agilent 1100 HPLC G1367A Autosampler

Ref : 2340992-18-CP
Condition : Used
Manufacturer : Agilent
Model : 1100 HPLC G1367A
Short Description : Autosampler
Year(s) : -
Quantity : 1
Location : Seller or machines location:
AMERICA North (USA-Canada-Mexico)
Last check : 15 Jul. 2022

Specifications :
GLP Features : Early maintainece feedback (EMF), electronic records of maintainence and errors
Communications : Controller-area network (CAN). RS232C, APG remote standard, optional four external contact closures and BCD vial number output
Safety Features : Leak detection and safe leak handling, low voltages in maintence areas, error detection and display
Injection Range : 0.1 – 100 µl in 0.1 µl increments Up to 1500 µl with multiple draw (hardware modification required)
Precision : Typically < 0.5 % RSD of peak areas from 5 – 100 µl, Typically < 1 % RSD from 1 – 5 µl
Sample Viscosity Range : 0.2 – 5 cp
Sample Capacity : 2 × well-plates (MTP) + 10 × 2 ml vials 100 x 2 ml in one tray 40 x 2 ml in half tray
Injection Cycle Time : Typically < 30 s using
Following standard conditions:
Default draw speed: 200 µl/min
Default eject speed: 200 µl/min
Injection volume: 5 µl

Carry-Over :
Typically < 0.01 % using the following conditions: Column: 125 x 4 mm Hypersil ODS, 5 µm
Mobile phase: Water/Acetonitrile = 80/20
Flow rate: 1 ml/min
Injection volume: 1 µl caffeine (1 mg/ml), 5 µl water to test carryover
Outside wash of needle before injection: 20 sec with water using flush port