Agilent 8453 G1103A, UV VIS Spectrophotometer
AMERICA North (USA-Canada-Mexico)
Wavelength range 190 – 1100 nm
EP resolution test > 1.6 spectrum of a 0.02% v/v solution oftoluene in hexane, ratio abs. at
Stray light < 0.03 % at 340 nm (NaNO2, ASTM)
< 0.05% at 220 nm (NaI, ASTM)
< 1% at 200 nm (KCl, EP)
Wavelength accuracy < ±0.5 nm (NIST 2034)*
< ±0.2 nm at 486.0 and 656.1nm
Wavelength reproducibility < ±0.02 nm ten consecutive scans (NIST 2034)
Photometric accuracy < ±0.005 A at 1 A (NIST 930e)
od
Photometric noise < 0.0002 A sixty 0.5-s scans at 0 A, 500 nm, rms
Photometric stability < 0.001 A/h at 0 A, 340 nm, after 1-hour warm-up, measured over 1 hour, every 5 seconds, constant ambient temperature
Baseline flatness < 0.001 A 0.5-second blank, 0.5-second scan, rms
Typical scan time 1.5 second full range
Slit width 1nm
WEIGHT 40 lbs
DIMENSIONS 26 × 18 × 15 in
INCLUDES Power Cord
VOLTAGE 120V/220V
HERTZ 50Hz/60Hz