Menu

Ziehm Quantum

Ref : 2524156-14-CP
Condition : Used
Manufacturer : Ziehm
Model : Quantum
Year(s) : 2008
Quantity : 1
Location : Seller or machines location:
AMERICA North (USA-Canada-Mexico)
Last check : 02 Oct. 2024

Dimensions
Orbital rotation: 135° a
Angulation: ± 225°
Swiveling (panning): ±10°
Horizontal movement: 220 mm
Vertical movement: 430 mm

C-Arm
Source–image receptor distance: 970 mm
Source-image receptor Opt 31: 1007mm
Vertical free space: 750 mm
Vertical free space: 787.4mm
Immersion depth: 680 mm

Image Intensifier
Nominal sizes: 23 / 17 / opt (10) cm
Max. Operating Data
Fluoroscopy Generator: 110 kV / 6 mA - 80 kV / 6 mA
Direct Radiography: 110 kV / 8 mA - 80 kV/8 mA
Radiography: 110 kV / 18 mA, 110 kV / 20 mA - 80 kV / 18mA, 80 kV / 20 mA

Max Power Output
Fluoroscopy: 660W (110 kv / 6mA)
Direct Radiography: 880 W (110 kV / 8 mA)
Direct Radiography: 1980W (110 kV / 18 mA), 2200W (110kV / 20mA)

Nominal Electric Power
2000W at 100kV / 20 mA / 0.1s

X-Ray Tube
Dual-focus stationary-anode tube

Focal Point Nominal Size
Nagel:
Small - 0.6 acc. to IEC
Large 1.5 acc to IEC

Toshiba
Small - 0.5 acc. to IEC
Large - 1.5 acc to IEC

Power Specifications
Power: 120 Vac 60 Hz Option:230 Vac 60 Hz
Direct Radiography: 40 - 110 kV
Tube Current: 2 mA min/ 20mA max
mAs Range: .4 mAs min./100mAs max
Fluoroscopy: 40-110kV - 0.1-6mA
Pulsed mode: Pluse Max 3F/s
Digital Radiography: 40-110 kV
Operating Frequency: 20 kHz

Environmental
During Storage Conditions
Temperature: -10°C to +60°C
Relative air humidity: 10% - 95% non Condensing max.
During Operation Environmental
Temperature: +10°C to +35°C
Relative air humidity: 35%-75% in Condensing max.