Drager Zeus SW 3.08
EUROPE (Western and Northern)
Breathing system with APL valve
2 DIVA dosing modules for sevoflurane and desflurane
Drägersorb CLIC replacement soda lime - prevents skin contact
DRÄGER Infinity C500 touchscreen monitor
Dräger VentStar® Anesthesia Breathing Circuits
With swivel arm
Technical Details:
Gas delivery
Modes of operation Fresh-gas control, auto control with insp. O2 control
Fresh-gas flow 0.25 – 18 l/min (fresh-gas control);0 – 18 l/min (auto control);
closed system mode (uptake)
O2 control 21 – 100 % (in fresh-gas control with AIR); Oxygen Ratio Controller
Carrier gases N2O and Air
O2 flush > 35 l/min
O2 safety flow 0 – 12 l/min
Auxiliary O2 flow tube 0 – 16 l/min (e.g. for regional anaesthesia)
Anaesthetic agent delivery DIVA (Direct Injection of Volatile Agents):
Mode of operation of DIVA modules Fresh-gas control, auto control with
expiratory agent control (2 plug-in slots for ISO, SEV, DES)
Max. filling volume 315 ml
Delivery range: Fresh-gas control (fresh-gas settings) ISO: 0 – 5 kPa;
SEV: 0 – 8 kPa; DES: 0 – 18 kPa
Delivery range: Auto control (expiratory agent settings) ISO: 0 – 2.5 kPa;
SEV: 0 – 5 kPa; DES: 0 – 12 kPa