Waters 186015001 Acquity Binary Solvent Manager
AMERICA North (USA-Canada-Mexico)
Number of Solvents Up to four
Solvent Conditioning Vacuum degassing (six-channel): one channel per solvent, and two channels for Sample Manager wash solvent
Operating Flow Range Rate 0.010 to 2.000 mL/min, in 0.001 mL increments
Compressibility Compensation Automatic and continuous
Effective System Delay Volume < 120 μL, independent of system backpressure
(with standard 50-μL mixer installed)
Plunger Seal Wash Integral, active, programmable
Gradient Profiles Eleven gradient curves [including linear, step (2), concave (4), and convex (4)]
Wet Prime Automatic
Maximum Operating Pressure 15,000 psi up to 1 mL/min, 9,000 psi up to 2 mL/min per pump,
not more than 15,000 psi total
Composition Accuracy ±0.5% absolute (full scale) from 5 to 95% of flow rates
from 0.5 to 2.0 mL/min (Contact Waters for conditions used)
Composition Precision 0.15% RSD or ±0.04 min SD, whichever is greater, based on retention time
Flow Precision 0.075% RSD or ±0.02 min SD, six replicates, based on retention time
or volumetric measures (0.500 to 2.000 mL/min)
Flow Accuracy ±1.0% at 0.5 mL/min with degassed methanol, per Waters AQT/SystemsQT protocol
Primary Wetted Materials 316 stainless steel, UHMWPE, sapphire, ruby, FEP, PTFE, ETFE,
diamond-like coating, PEEK and PEEK alloys, titanium alloys
WEIGHT 80 lbs
DIMENSIONS 34 × 22 × 18 in
INCLUDES Power Cord
VOLTAGE 120V/220V
HERTZ 50Hz/60Hz